Order #: 9103-A-156-TRAN
Mfg #: 2431947
Fluke Calibration 9103-A Field Dry-Well Calibrator with Insert A (3103-2) 110V, 50/60Hz
Order #: 9103-A-156-TRAN
Mfg #: 2431947
Product Highlights
The 9103 covers below-ambient temperatures as low as –25 °C. The 9103 is stable to ±0.02 °C, and its display is calibrated to an accuracy of ±0.25 °C at all temperatures within its range. In just eight minutes, 0 °C is reached, and 100 °C is reached in six minutes, so your time is spent calibrating—not waiting.
The 9103 reaches temperatures 50 °C below ambient, so –25 °C is reached under normal ambient conditions. Our competitors like to advertise their units as reaching –45 °C when they really mean –45 °C below ambient, which typically means it will go to –20 °C. Our unit does not require you to work in a walk-in freezer to achieve its full advertised range.
Choose one of three removable inserts sized for probes from 1/16 inch to 1/2 inch in diameter. Insert A handles a full range of probe sizes with a single well of each size. Insert B features two wells each of 3/8, 1/4, and 3/16 inches in diameter for doing comparison calibrations. Insert C has six 1/4-inch-diameter wells for multiple probe calibrations, and Insert D has three pairs of metric sized wells.
- Range: –25°C to 140°C (–13°F to 284°F) at 23°C ambient
- Accuracy: ±0.25°C
- Stability: ±0.02°C at –25°C ±0.04°C at 140°C
- Well-to-Well Uniformity: ±0.1°C between similarly sized wells
- Heating Times: 18 minutes from ambient to 140°C
- Cooling Times: 20 minutes from ambient to –25°C
- Stabilization Time: 7 minutes
- Immersion Depth: 124mm (4.875in)
- Size (W x H x D): 143 x 261 x 245mm (5.63 x 10.25 x 9.63in)
- Weight: 12 lb (5.7 kg)
- NIST-Traceable Certificate Data at: –25°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 140°C
Each Unit Includes
- Nist Traceable Certificate
- Insert A
Webinars
Automating and Documenting Temperature Calibration Webinar with Fluke Calibration