Product Highlights
The Test Equity 1007C Temperature Chamber allows up to 256 steps to be programmed into as many as 40 namable profiles. The context sensitive information key and guided steps make profile programming fast and easy. A four line backlit LCD displays programming, setup, operating and help information. A large LED readout indicates the actual chamber temperature with 0.1° resolution. Internal logic provides refrigeration compressor control for responsive and reliable performance. The Test Equity 1007C includes two alarms and seven event outputs to control remote devices. RS-232C communications is also included. GPIB, Ethernet and analog retransmit are available as options.
Features and specifications of the 1007C Environmental Chamber include:
- -73C to +175C Temperature Range (opt. to +205C)
- 7 Cu Ft Workspace, 24 W x 21 H x 24 D (198 Liters)
- Programmable Temperature Controller
- RS-232 Interface (Opt. GPIB and Ethernet)
- High/Low Limit Control and Alarm
- Viewing Window & Interior Light
- 4" Access Ports on Left & Right Side
- Non-CFC Cascade Refrigeration
- ETL Listed to UL 61010-1 and CSA C22.2 No. 61010-1
- CE Mark (with export voltage option 0725)
- Made in the USA
- Frame Type: Floor-Standing
- Input Voltage: 208/230V, 60Hz, 1/3 PH
- Features: Window/Light
- Interface: RS-232, Ethernet
- LN2 Boost Cooling: No
- Maximum AC Current: 19A to 28A
- No. of Ports: 2 x 4"
No-Sweat Cabinet
Testequity uses a combination of structural foam and fiberglass blanket insulation for a higher combined R-factor than just fiberglass alone. The door sealing surface is kept warm by circulating excess heat from the refrigeration system. The result is a cabinet that does not sweat due to condensation during low temperature operation.
The window is constructed of 6 panes of tempered glass separated by low-conductivity spacers to keep thermal losses to a minimum. Our unique design virtually eliminates exterior window sweating during normal temperature cycling conditions.
The stainless-steel interior liner is heliarc welded, while the refrigeration compartment behind the liner is double-insulated and carefully sealed. These design features reduce thermal losses and prevent moisture migration, ensuring years of optimum performance while keeping the cabinet rust-free.
Reliability & Safety That's Designed In
Non-CFC refrigerants are used in a cascade (two compressors) configuration. Expansion valves provide optimum performance throughout the operating range. Pressure regulators keep the operating pressures within safe limits regardless of load conditions. Liquid injection ensures cool compressor operation during high-temperature pull down for long life. Sequential starting of each compressor reduces the current demand on start-up. Each compressor is protected by high/low pressure switches. Care has been taken to use a minimum number of fittings, joints and welds to reduce the possibility of refrigerant leaks.
All electrical control components are UL approved. Each branch circuit is individually protected with class-CC fuses. A main power disconnect switch is provided on the front panel. The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.
The side and rear panels are removable to permit complete access to the refrigeration system for ease of service and preventive maintenance. Shown above with the optional circular chart recorder, which mounts conveniently in the front.