Keysight E36154A DC power supply, single-output, auto-range: 30V, 80A, 800W
DC power supply, single-output, auto-range: 30V, 80A, 800W
Order #: E36154A
Mfg #: E36154A
This product is available for shipping to the United States and Canada only.
Product Highlights
Features
- Plenty of usable power
- Autoranging
- Peak power handling up to 3X max power rating
- Safe, clean, and reliable power
- Low output ripple and noise
- Excellent line/load regulation
- 2-wire or 4-wire remote sense
- Detachable high current front binding post
- Over-voltage, over-current, and over-temperature protection
- Convenient benchtop capabilities and intuitive interfaces
- Thermal-control fan speed for minimal acoustic noise
- 4-wire front output terminal including sense and ground
- 4.3-inch LCD color display
- Individual knobs for voltage and current
- LAN/LXI, USB and GPIB (upgrade option) interfaces
- Advanced characterization
- Built-in voltage and current measurements
- Data logging
- Output sequencing and syncing with digital I/O
- LIST mode programming
- Low range current measurement
- Adjustable voltage slew rate
- Scope View (upgrade option)
- Arbitrary waveform generator (AWG) (upgrade option)
- Application Software and Automation
- BV0003B Pathwave BenchVue Power Supply App
- BV9200B/BV9201B BenchVue Advanced Power Control and Analysis
-
$704.00
-
$522.00
Model |
E36154A |
E36155A |
Power Output |
800W |
800W |
No. of Channel |
1 |
1 |
DC Output Rating (0 to 40 °C) |
0 to 30 V |
0 to 60 V |
0 to 80 A |
0 to 40 A |
Load Regulation ± (% of output + offset) |
Voltage |
< 0.01% + 2 mV |
Current |
< 0.1% + 2 mA |
Line Regulation ± (% of output + offset) |
Voltage |
< 0.01% + 2 mV |
Current |
< 0.1% + 2 mA |
Output Ripple and Noise (at approximately 23 °C) |
Normal mode voltage, Vpp (20 Hz to 20 MHz) |
< 75 mV |
Normal mode voltage, Vrms (20 Hz to 10 MHz) |
< 5 mV |
Programming Accuracy ± (% of output + offset) at 23 °C ± 5 °C for 12 months |
Voltage |
0.03% + 6 mV |
0.03% + 10 mV |
Current |
0.1% + 20 mA |
0.1% + 10 mA |
Readback Accuracy ± (% of output + offset) at 23 °C ± 5 °C for 12 months |
Voltage |
0.04% + 6 mV |
0.04% + 10 mV |
Current |
0.1% + 20 mA |
0.1% + 10 mA |
Low Range Current |
0.1% + 5 mA |
0.1% + 4 mA |
Load Transient Recovery Time |
Voltage Settling Band |
75 mV |
150 mV |
Time |
< 1 ms |