IS Press Indicator 3.5bar/50psi 350kPa Abs G1/8 BSP Female
Order #: DPI705EIS-1-08A-P1-H1-U0-OP0
Mfg #: DPI705EIS-1-08A-P1-H1-U0-OP0
Druck DPI705EIS IS Pressure Indicator, 3.5bar/50psi ,350kPa, Absolute
IS Press Indicator 3.5bar/50psi 350kPa Abs G1/8 BSP Female
Order #: DPI705EIS-1-08A-P1-H1-U0-OP0
Mfg #: DPI705EIS-1-08A-P1-H1-U0-OP0
Product Highlights
- Unit is configured with:
- Model Type: DPI705EIS : Hazardous Area Pressure Indicator
- Accuracy: 1 : Standard
- Pressure Range and Reference Type : 08A : 3.5 bar / 50 psi / 350 kPa
- Pressure Fitting : P1 : G1/8 BSP Female
- Hazardous Area Approvals : H1 : ATEX, IECEx, CSA, CCOE, XPL, KCS, NEPSI, ECASEx
- Pressure Units : U0 : All Pressure Units
- Options : OP0 : No Option Required
- Intrinsically Safe version for hazardous areas
- Ranges 0 to 25 mbar through to 700 bar (0 to 10 inH2O through to 10,000 psi) with internal sensor
- Optional external sensor 0 to 25 mbar through to 1400 bar (0 to 10 inH2O through to 20,000 psi)
- 48 total pressure ranges
- Gauge, Absolute, and Differential choices
- 19 programmable pressure units such as psi, inH2O, bar, inHg
- Optional negative calibration for vacuum
- Impact resistant, Rugged, lightweight handheld design
- Leak test, tare, min/max, alarm, and filter
- Optional remote RTD temperature sensors
Druck DPI 705E-IS Intrinsically Safe Pressure Indicator with Internal Sensor
- Intrinsically Safe version for hazardous areas
- Ranges 0 to 25 mbar through to 700 bar (0 to 10 inH2O through to 10,000 psi) with internal sensor
- Optional external sensor 0 to 25 mbar through to 1400 bar (0 to 10 inH2O through to 20,000 psi)
DPI 705E Series Druck Handheld Pressure Indicators
The Druck DPI 705E Series of handheld pressure and optional temperature indicators combine tough and rugged design with accurate and reliable measurements. Compact and robust, the DPI 705E Series is designed for single handed operation and provides many essential features required for routine maintenance and system troubleshooting.
DPI 705E Improves Test Efficiency
- Power on and use. No warm up time required.
- 19 pressure measurement units (See below)
- 1, 3 or 5 minute leak test
- Impact resistant, sealed to IP54
- Large digit high-contrast LCD with backlight
- Optional carrying case with belt loop
- Optional magnetic/loop hanging strap
- Long battery life
- Integral desk stand and hanger
Special Features Pressure Units
- mbar, bar, Pa, hPa, Pa, MPa, psi , lb/ft2, kgf/cm2, kgf/m2, mmHg(0°C), mHg(0°C), inHg 0, mmH2O, cmH2O, mH2O, inH2O (4°C, 20°C), ftH2O (4°C, 20°C)
Calibration Certificates
- Supplied as standard with bar, psi and kPa
- Optional UKAS approved calibrations available
Zero
Zero pressure correction (gauge/differential sensors)
Tare
0 to 100% FS temporary zero offset capability by subtracting the current reading from subsequent measurements.
Filter
Enables a filtered pressure reading by showing a rolling average of the last 10 measurements. Providing a more stable reading of a noisy measurement.
Alarm
User adjustable; high and low pressure alarms with visual (Bell Icon, Pressure Reading and Backlight flash) and audible warning for 60 seconds.
Calibration
Used in conjunction with the optional pneumatic or hydraulic hand pumps, the DPI 705E Series provides a simple, low-cost calibration solution.
- Druck IO705E-CASE-IS DPI705E Carry Case (IS)DPI705E Carry Case (IS)
Order #: IO705E-CASE-IS
Mfg #: IO705E-CASE-IS
Availability: Ships in 3-4 weeksSpecial Price $63.49 Regular Price $69.02 - Druck IO705E-STRAP DPI705E Hanging StrapDPI705E Hanging Strap
Order #: IO705E-STRAP
Mfg #: IO705E-STRAP
Availability:In stockSpecial Price $63.49 Regular Price $69.02 - Druck IO-ADAPT-QF G1/8M to Quick-fit AdaptorG1/8M to Quick-fit Adaptor
Order #: IO-ADAPT-QF
Mfg #: IO-ADAPT-QF
Availability: Ships in 3-4 weeksSpecial Price $99.69 Regular Price $108.36
General Specifications | |
Working Pressure | 110% FS (alarm active beyond this range) |
Sealing | IP54 |
Operating temperature | -10 to 50°C (14 to 122°F) |
Storage temperature | -20 to 70°C (-4 to 158°F) |
Humidity | 0 to 90% RH non condensing |
Shock and vibration | MIL-PRF-28800F for Class II equipment.1 m Drop Tested @ -20°C (-4°F) |
EMC | BS EN 61326-1 |
Electrical safety | BS EN 61010-1, UL 61010-1 |
Pressure safety | Pressure equipment directive class SEP. UL61010, Safe to 2xFS |
Approval | CE marked, RCM |
Size and weight | L 200 mm, W 95 mm, D 43 mm, 563 g maximum |
RoHS | Compliant |
Media Compatibility | Media to be compatible with Stainless Steel and Hastelloy |
Resolution | 5 digits |