Calibration Service Levels
Depending on the quality policy requirements of your company and the required industry standards, Transcat offers a choice of non-accredited and 17025 accredited calibration service levels. Click here for a detailed description of Transcat's new calibration service levels.
- Calibration Without Data (Non-accredited)
- Calibration With Data (Non-accredited)
- Accredited Calibration With Data & Uncertainties (Simple Acceptance)
- With Optional Guard Banding
- With Optional Probability of Compliance to the Specifications (PCS)
- With Optional Guard Banding and PCS
No matter which calibration service option you choose, when you calibrate with Transcat, you can expect prompt and responsive service including five to seven business day turnaround on most instruments.
Note: Per ILAC P14, Calibrations that do not include reported measurement uncertainties prevent the respective instrument from propagating traceability. Visit our resource page for more information on ILAC P14, Z540.3 and other quality requirements.
ISO 9001:2015 Compliant Quality System | |
ISO 17025:2017 Compliant Quality System | |
21 CFR Part 820.75 Validated Metrology Management System | |
10 CFR Part 21 and Part 50 Appendix B compliant | |
Calibrated to manufacturer's specifications or customer requirements |
ISO 9001:2015 Compliant Quality System | |
ISO 17025:2017 Compliant Quality System | |
21 CFR Part 820.75 Validated Metrology Management System | |
10 CFR Part 21 and Part 50 Appendix B compliant | |
Calibrated to manufacturer's specifications or customer requirements | |
As-found/As-left data provided | |
Tolerances provided |
ISO 9001:2015 Compliant Quality System | |
ISO/IEC 17025:2017 Accredited | |
21 CFR Part 820.75 Validated Metrology Management System | |
10 CFR Part 21 and Part 50 Appendix B compliant | |
Calibrated to manufacturer's specifications or customer requirements | |
As-found/As-left data provided | |
Tolerances provided | |
NQA-1 Compliant | |
Z540.1 Compliant | |
Accreditation includes ANSI/NCSL Z540.3-2006 1 | |
Traceable to SI Units | |
Test Uncertainty Ratio (TUR) calculation provided | |
Enables analysis of an instrument's impact on a process | |
Detailed uncertainties provided | |
Enables evaluation of risk in instrument applications | |
View Sample Cert
|
|
Optional | |
Measurement of Guard Banding 2 ) | |
Transcat's Exclusive Probability of Compliance to the Specifications (PCS) Service 3, 4 ) |
View PCS Cert
View GB/PCS Cert
1 - Read our FAQ on ANSI Z540.3.
2 - Read our white paper on Guard Banding 101.
3 - Download our white paper on Transcat’s Exclusive Probability of Compliance to the Specification (PCS) Service.
4 - Optional PCS service may be added to this service level at an optional fee. The customer must specify the decision rule to be applied to the Guard Banding.
Calibration Service Levels
Depending on the quality policy requirements of your company and the required industry standards, Transcat offers a choice of non-accredited and 17025 accredited calibration service levels.
- Certificate Without Data
- Certificate With Before and After Data
- Accredited Certificate With Data & Uncertainties (Simple Acceptance)
- Accredited Certificate With Data & Uncertainties & Guard Banding
- Accredited Certificate With Data & Uncertainties & PCS
- Accredited Certificate With Data & Uncertainties & Guard Banding & PCS
No matter which calibration service option you choose, when you calibrate with Transcat, you can expect prompt and responsive service including five to seven business day turnaround on most instruments.
Note: Per ILAC P14, Calibrations that do not include reported measurement uncertainties prevent the respective instrument from propagating traceability. Visit our resource page for more information on ILAC P14, Z540.3 and other quality requirements.
Calibration Without Data Sample |
Calibration With Data Sample |
Calibration With Data & Uncertainties Sample |
|
---|---|---|---|
ISO 9001:2015 Compliant Quality System | |||
21 CFR Part 820.75 Validated Metrology Management System | |||
10 CFR Part 21 and Part 50 Appendix B compliant | |||
Calibrated to manufacturer's specifications or customer requirements | |||
As-found/As-left data provided | |||
Tolerances provided | |||
NQA-1 Compliant | |||
Z540.1 Compliant | |||
Accreditation includes ANSI/NCSL Z540.3-20062 | |||
Traceable to SI Units | |||
Test Uncertainty Ratio (TUR) calculation provided | |||
ISO/IEC 17025:2017 Accredited | |||
Enables analysis of an instrument's impact on a process | |||
Detailed uncertainties provided | |||
Enables evaluation of risk in instrument applications | |||
Measurement of Guard Banding3 | Optional 1 | ||
Transcat's Exclusive Probability of Compliance to the Specifications (PCS) Service4 | Optional 1 | ||
1 - Optional Guard Banding and/or Transcat's PCS service may be added to this service level at an optional fee. 2 - Read our FAQ on ANSI Z540.3. 3 - Read our white paper on Guard Banding 101. 4 - Download our white paper on Transcat’s Exclusive Probability of Compliance to the Specification (PCS) Service. |
For more information about our accreditation or your calibration requirements, contact one of our calibration professionals at 800-828-1470 or contact us.
Detailed Description of Calibration Service Levels
Transcat offers several service levels to accommodate the varying quality system requirements a customer may have regarding an instrument’s calibration. We understand that one size does not fit all. Considering this, you have your choice from the service options below. You can view a summary of these service levels and sample certificates here.
Transcat’s calibration specialists are happy to help answer your questions, however the ultimate decision on which calibration level that is required is solely at your discretion. All Transcat calibration reports are available online via Transcat’s C3 portal.
1.
Calibration without Data (Non-accredited)
This level of service provides you with a calibration certificate and sticker for your instrument. We will evaluate it to a Pass/Fail criterion and no as-found or as-left data will be recorded or provided. This level is unaccredited and not traceable to a national standard. This level of calibration will not have an accreditation logo.
2.
Calibration with Data (Non-accredited)
This level of service provides you with a calibration in which the as found and as left data will be recorded and provided along with your calibration certificate and sticker for your instrument. We will also provide you a separate out of tolerance (OOT) report, as well as Test Uncertainty Ratios (TUR) when the TUR’s are < 4:1. This level is unaccredited and not traceable to a national standard. This level of calibration will not have an accreditation logo.
3.
Accredited Calibration With Data & Uncertainties (Simple Acceptance)
This level of service provides you with a calibration in which the as found and as left data with measurement uncertainties and test uncertainty ratios (TUR) will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025, and meets NIST Traceability requirements. Transcat’s default Guard Banding method will be Simple Acceptance (ASME B89.7.3.1) to be applied as your Decision Rule unless you select one of the Guard Band methods below. This level of calibration will have an accreditation logo.
a.
With Optional Guard Banding
This level of service provides you with a calibration in which the as-found and as left data with measurement uncertainties, test uncertainty ratios (TUR) and Guard Banding will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards and meets NIST Traceability requirements. This level of calibration will have an accreditation logo.
Customers may also choose an alternate Guard Banding Decision Rule. This must be specified at the time of placing their order
- ANSI/NCSL Z540.3: 2006, Method 5
- ANSI/NCSL Z540.3: 2006 Method 6
- ILAC G8:2019, Table 1
- ASME B89.7.3.1, clause 5.2, Stringent Acceptance
- ASME B89.7.3.1, clause 5.3, Relaxed Acceptance
b.
With Optional Probability of Compliance to the Specifications (PCS)
This level of service provides you with a calibration in which the as-found and as left data with full test uncertainty ratios (TUR) and Transcat’s PCS measurement risk feature will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards, and meets NIST Traceability requirements. Transcat’s PCS feature quantifies the measurement risk with a threshold level of 98% PCS (equivalent to 2% PFA); you may set this threshold to a different level, depending on your quality requirements and risk management policy. For more information on Transcat’s PCS tool, check out our white paper. This level of calibration will have an accreditation logo.
c.
With Optional Guard Banding and PCS
This level of service provides you with a calibration in which the as-found and as left data with measurement uncertainties, test uncertainty ratios (TUR) and Guard Banding and Transcat’s PCS feature will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards, and meets NIST Traceability requirements. With this service level, the Guard Band method selected will be used to identify Pass/Fail for the Guard Band Acceptance Limits while Transcat’s PCS feature will simply quantify the exact measurement risk that exists for each measurement result. This level of calibration will have an accreditation logo.
Customers may also choose an alternate Guard Banding Decision Rule. This must be specified at the time of placing their order
- ANSI/NCSL Z540.3: 2006, Method 5
- ANSI/NCSL Z540.3: 2006 Method 6
- ILAC G8:2019, Table 1
- ASME B89.7.3.1, clause 5.2, Stringent Acceptance
- ASME B89.7.3.1, clause 5.3, Relaxed Acceptance
Transcat offers several service levels to accommodate the varying quality system requirements a customer may have regarding calibration. Transcat’s calibration specialists are happy to help answer your questions, however the ultimate decision on which calibration is required is solely at the customer’s discretion. All Transcat calibration reports are available online via Transcat’s C3 portal.
1.
Level 1 - Unaccredited Calibration without Data
This level of service provides you with a calibration certificate and sticker for your instrument. We will evaluate it to a Pass/Fail criterion and no as-found or as-left data will be recorded or provided. This level is unaccredited and not traceable to a national standard. This level of calibration will not have an accreditation logo.
2.
Level 2 - Unaccredited Calibration with Data
This level of service provides you with a calibration in which the as found and as left data will be recorded and provided along with your calibration certificate and sticker for your instrument. We will also provide you a separate out of tolerance (OOT) report, as well as Test Accuracy Ratios (TAR) when the TAR’s are <4:1. This level is unaccredited and not traceable to a national standard. This level of calibration will not have an accreditation logo.
3.
Level 3 - Accredited/Traceable Calibration with Data and Uncertainties
This level of service provides you with a calibration in which the as found and as left data with full test uncertainty ratios (TUR) will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards, and meeting NIST Traceability requirements. This level of calibration will have an accreditation logo.
4.
Level 4 - Accredited/Traceable Calibration with Data and Uncertainties and Guard Banding
This level of service provides you with a calibration in which the as-found and as left data with full test uncertainty ratios (TUR) and Guard Banding will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards and meeting NIST Traceability requirements. Transcat’s default Guard Banding method will be Simple Acceptance (ASME B89.7.3.1). This level of calibration will have an accreditation logo.
Customers may also choose an alternate Guard Banding Decision Rule. This must be specified at the time of placing their order
- ANSI/NCSL Z540.3 Method 5
- ANSI/NCSL Z540.3 Method 6
- ILAC G8
- ISO 14253-1
5.
Level 5 - Accredited/Traceable Calibration with Data and Uncertainties with Transcat's PCS (Probability of Compliance to the Specification) Feature
This level of service provides you with a calibration in which the as-found and as left data with full test uncertainty ratios (TUR) and Transcat’s PCS feature will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards, and meeting NIST Traceability requirements. Transcat’s PCS feature will help normalize your data and help you determine if measurement risk exists. For more information on Transcat’s PCS tool, check out our white paper. This level of calibration will have an accreditation logo.
6.
Level 6 - Accredited/Traceable Calibration with Data and Uncertainties and Guard Banding and Transcat's PCS (Probability of Compliance to the Specification) Feature
This level of service provides you with a calibration in which the as-found and as left data with full test uncertainty ratios (TUR) and Guard Banding and Transcat’s PCS feature will be recorded and provided along with your calibration certificate and sticker for your instrument. This level fully complies with ISO 17025 and Z540.3 standards, and meeting NIST Traceability requirements. Transcat’s default Guard Banding method will be Simple Acceptance (ASME B89.7.3.1). In addition, Transcat’s PCS feature will help normalize your data and help you determine if measurement risk exists. This level of calibration will have an accreditation logo.
Customers may also choose an alternate Guard Banding Decision Rule. This must be specified at the time of placing their order
- ANSI/NCSL Z540.3 Method 5
- ANSI/NCSL Z540.3 Method 6
- ILAC G8
- ISO 14253-1
Calibration Service Levels
Depending on the quality policy requirements of your company and the required industry standards, Transcat offers a choice of 17025 accredited calibration service levels:
- Accredited Certificate Without Data
- Accredited Certificate With Before and After Data
- Accredited Certificate With Data & Uncertainties
No matter which calibration service option you choose, when you calibrate with Transcat, you can expect prompt and responsive service including five to seven business day turnaround on most instruments.
Note: Per ILAC P14, Calibrations that do not include reported measurement uncertainties prevent the respective instrument from propagating traceability.
Certificate Without Data Sample | Certificate With Data Sample | Certificate With Data & Uncertainties Sample | |
---|---|---|---|
ISO/IEC 17025 Accredited | |||
Calibrated to manufacturer's specifications or customer requirements | |||
10 CFR Part 21 and Part 50 Appendix B compliant | |||
21 CFR Part 820.75 Validated Metrology Management System | |||
Greater than 4:1 Test Uncertainty Ratio (with exceptions noted on your certificate) |
|||
Accreditation includes ANSI/NCSL Z540-1-1994 | |||
ISO 9001 Compliant Quality System | |||
Traceable to SI Units | |||
As-found/As-left data provided | |||
Tolerances provided | |||
Enables analysis of an instrument's impact on a process | |||
Detailed uncertainties provided | |||
Test Uncertainty Ratio (TUR) calculation provided | |||
Enables evaluation of risk in instrument applications ANSI/NCSL Z540.3 |
For more information about our accreditation or your calibration requirements, contact one of our calibration professionals at 800-828-1470 or contact us.