The store will not work correctly when cookies are disabled.
JavaScript seems to be disabled in your browser.
For the best experience on our site, be sure to turn on Javascript in your browser.
Your company account is blocked and you cannot place orders. If you have questions, please contact your company administrator.
Fluke 714B Thermocouple Calibrator
Thermocouple Calibrator
Order #: 714B
Mfg #: 4387869
Availability: Ships in 1-2 weeks
This product is available for shipping to the United States, Canada, and Puerto Rico only.
This item is available as RENTAL at Call for price
Shop now
Overview
For the temperature calibration professional that wants a highly accurate, easy-to-use, single function thermocouple temperature calibrator the Fluke 714B is the ideal test tool.
Product Highlights The 714B can measure and simulate (17) different thermocouple types and millivolts Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal Hanging tool designed in and included with every unit Configurable 0% and 100% source settings for quick 25% linearity checks Linear ramp and 25% step auto ramp based on 0% and 100% settings Dual inputs and backlit display for easy interpretation of measurements Power down settings remembered at power up for easy restart of tests 1-year and 2-year specifications and traceable certificate of calibration Each Unit Includes Magnetic Hanging Tool Batteries Manual Traceable Calibration Certificate Test Leads
Special Price
$335.16
Regular Price
$376.59
Special Price
$265.06
Regular Price
$297.83
Special Price
$69.76
Regular Price
$78.39
General Specifications
Maximum voltage applied between any terminal and earth ground or between any two terminals
30 V
Operating temperature
-10°C to 50°C
Storage temperature
-30°C to 60°C
Operating altitude
2,000 m
Storage altitude
12,000 m
Relative humidity (% RH operating without condensation)
Non condensing
90% (10°C to 30°C)
75% (30°C to 40°C)
45% (40°C to 50°C)
(Without condensation)
Vibration requirements
MIL-T-28800E, Class 2
Drop text requirements
1 m
IP Rating
IEC 60529: IP52
Electromagnetic environment
IEC 61326-1, Portable
Safety
IEC 61010-1, Max 30 V to earth, Pollution Degree 2
Power supply
4 AA NEDA 1.5A IEC LR6 batteries
Size (H x W x L)
52.5 x 84 x 188.5 mm
Weight
515 g
DC mA Measurement
Resolution
Range
Accuracy (% of reading + counts)
0-24 mA
0.001 mA
0.010% + 2 μA
Temperature coefficient
±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)
Millivolt Measurement and Source
Resolution
Range
Accuracy (% of reading +counts)
-10 mV to 75 mV
0.01 mV
0.015% + 10 μA
Temperature coefficient
±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C)
Thermocouple Input and Output
TC Type
Range (°C)
Measure (°C)
Source (°C)
1 year
2 year
1 year
2 year
E
-250 to 200°C
1.3
2
0.6
0.9
-200 to -100°C
0.5
0.8
0.3
0.4
-100 to 600°C
0.3
0.4
0.3
0.4
600 to 1000°C
0.4
0.6
0.2
0.3
N
-200 to -100°C
1.0
1.5
0.6
0.9
-100 to 900°C
0.5
0.8
0.5
0.8
900 to 1300°C
0.6
0.9
0.3
0.4
J
-210 to -100°C
0.6
0.9
0.3
0.4
-100 to 800°C
0.3
0.4
0.2
0.3
800 to 1200°C
0.5
0.8
0.3
0.3
K
-200 to -100°C
0.7
1
0.4
0.6
-100 to 400°C
0.3
0.4
0.3
0.4
400 to 1200°C
0.5
0.8
0.3
0.4
1200 to 1372°C
0.7
1
0.3
0.4
T
-250 to -200°C
1.7
2.5
0.9
1.4
-200 to 0°C
0.6
0.9
0.4
0.6
0 to 400°C
0.3
0.4
0.3
0.4
B
600 to 800°C
1.3
2
1
1.5
800 to 1000°C
1
1.5
0.8
1.2
1000 to 1820°C
0.9
1.3
0.8
1.2
R
-20 to 0°C
2.3
2.8
1.2
1.8
0 to 100°C
1.5
2.2
1.1
1.7
100 to 1767°C
1
1.5
0.9
1.4
S
-20 to 0°C
2.3
2.8
1.2
1.8
0 to 200°C
1.5
2.1
1.1
1.7
200 to 1400°C
0.9
1.4
0.9
1.4
1400 to 1767°C
1.1
1.7
1
1.5
C
0 to 800°C
0.6
0.9
0.6
0.9
800 to 1200°C
0.8
1.2
0.7
1
1200 to 1800°C
1.1
1.6
0.9
1.4
1800 to 2316°C
2
3
1.3
2
L
-200 to -100°C
0.6
0.9
0.3
0.4
-100 to 800°C
0.3
0.4
0.2
0.3
800 to 900°C
0.5
0.8
0.2
0.3
U
-200 to 0°C
0.6
0.9
0.4
0.6
0 to 600°C
0.3
0.4
0.3
0.4
BP
0 to 1000°C
1
1.5
0.4
0.6
1000 to 2000°C
1.6
2.4
0.6
0.9
2000 to 2500°C
2
3
0.8
1.2
XK
-200 to 300°C
0.2
0.3
0.2
0.5
300 to 800°C
0.4
0.6
0.3
0.6
G
100 to 300°C
1.6
2.4
1.2
1.8
300 to 1500°C
1
1.5
1
1.5
1500 to 2320°C
2
3
1.6
2.4
D
0 to 300°C
1.6
2.4
1.2
1.8
300 to 1500°C
1
1.5
1
1.5
1500 to 2315°C
2
3
1.6
2.4
P
0 to 1000°C
1.6
2.4
0.6
0.9
1000 to 1395°C
2
3
0.8
1.2
M
-50 to 100°C
1.0
1.5
0.4
0.6
100 to 1000°C
1.6
2.4
0.6
0.9
1000 to 1410°C
2
3
0.8
1.2
All products in this series