The 1586A Super-DAQ is the most accurate and flexible temperature data acquisition system on the market. It scans and records temperature, dc voltage, dc current, and resistance for up to 40 input channels and scan speeds as fast as 10 channels per second. The Super-DAQ can be configured for use as a multi-channel data logger in the factory or as a precision reference thermometer for benchtop sensor calibration in the lab.
Availability: Ships in 3-4 weeks

Choose Your Options:

Add Calibration

You have chosen to add the following Calibration Services to Your Order

Make sure your instrument is calibrated and ready to use the minute you open the box by adding our expert calibration service. Why Calibrate? For a complete descriptions of Transcat Service Levels, click here.

Already have this this item and need calibration? Request a calibration quote

This product is available for shipping to the United States, Canada, and Puerto Rico only.

Product Highlights

  • Measure thermocouples, PRTs, thermistors, dc voltage, dc current, and resistance
  • Best-in-class temperature measurement accuracy
    • PRTs ± 0.005 °C (using external DAQ-STAQ Multiplexer)
    • Thermocouples: ± 0.5 °C (using High-Capacity Module and internal CJC)
    • Thermistors: ± 0.002 °C
  • Input channels: Up to 40 isolated universal inputs
  • Flexible configuration: Internal High-Capacity Module and/or DAQ-STAQ Multiplexer
  • Selectable scan speed: Up to 10 channels per second
  • Four modes of operation: Scan, Monitor, Measure, Digital Multimeter (DMM)
  • Real-time color trending: Chart up to four channels simultaneously
  • 6 1/2 digit display resolution for dc voltage, dc current, and resistance
  • Current reversal minimizes thermoelectric effects, improving PRT/RTD measurement accuracy
  • Automated sensor calibration: Control Fluke Calibration temperature sources such as dry-wells or micro-baths for automated calibration routines
  • Data storage: Records up to 20 MB of data and setup files to internal non-volatile memory or to an external USB drive. Transfer data to a PC using the USB drive or LAN connection and view data in Microsoft® Excel
  • Data security: Administrator and user profiles for protecting settings and ensuring test traceability
  • Mx + B scaling and channel offset zero function
  • Alarms: Two independent, user-defined alarms for each channel indicate when an upper or lower range has been exceeded
  • Supported by TQSoft and TQAero thermal validation software for FDA 21 CFR Part 11 and AMS 2750 compliant data collection

Webinars

Automating and Documenting Temperature Calibration Webinar with Fluke Calibration

Win a Generac Generator when purchasing $750 or more from Transcat.Win a Generac Generator when purchasing $750 or more from Transcat.

General Specifications

Mains Voltage 100 V Setting: 90 V to 110 V
120 V Setting: 108 V to 132 V
220 V Setting: 198 V to 242 V
240 V Setting: 216 V to 264 V
Frequency 47 Hz to 440 Hz
Power Consumption  36 VA peak (24 W average)
Environment Temperature Operating: 0 °C to 50 °C
Full accuracy: 18 °C to 28 °C
Storage: -20 °C to 70 °C
Warm-up: 1 hour to full accuracy specifications
Relative Humidity (non-condensing) Operating:  0 °C to 30 °C <80 %;
                   30 °C to 50 °C <50 %
Storage: -20 °C to 70 °C <95 %
Altitude Operating: 2,000 m
Storage: 12,000 m
Vibration and Shock  Complies with MIL-PRF-28800F Class 3
Channel Capacity Total analog channels: 45
Voltage/resistance channels: 41
Current channels:  5
Digital I/O: 8 bits
Totalizer: 1
Alarm outputs: 6
Trigger input: 1
Input Protection  50 V all functions, terminals and ranges
Math Channels Number of channels: 20
Operations: sum, difference, multiply, divide, polynomial, power, square root, reciprocal, exponential, logarithm, absolute value, average, maximum, minimum
Triggers interval, external (trigger input), alarm, remote (bus), manual, automated test
Memory Scan data RAM: 75,000 readings with timestamp
Data/Setup flash memory: 20 MB
USB Host Port Connector type: Type A
Function: Memory
File system: FAT32
Memory capacity: 32 GB
USB Device Port Connector type: Type B
Class: Instrument
Function: Control and data transfer
Command protocol: SCPI
LAN Function: Control and data transfer
Network protocols: Ethernet 10/100, TCP/IP
Command protocol: SCPI
RS-232 Connector: D-sub 9 pin (DE-9)
Baud rates: 1200, 2400, 4800, 9600, 19200, 38400
Function: Temperature source control output
Dimensions Height: 150 mm
Width: 245 mm
Depth: 385 mm Weight: 6 kg (typical configuration)
Shipping Weight: 9.5 kg (typical configuration)
Conformity CE, CSA, IEC 61010 3rd ed.