Product Highlights
- Dual-range pressure controller (water version)
- Interchangeable pressure module (sold separately)
- Pressure range: 0 to 15,000 psi (1,000 Bar)
- Accuracy of 0.02%FS, 0.01%FS, or 0.01%RD
- Dual range from -15 - 3000 psi (-1 - 200 Bar) to 0 - 15000 psi (0 - 1000) Bar
- Control stability 0.005%FS
- Ultra-High speed pressure generation and control
- No external pressure source required
- Control using internal or external pressure modules
- LAN, USB, RS232, and Ethernet communication
- Pressure modules are not included
Each Unit Includes
- 1 pc AC power cord (10A 250V)
- ISO17025 accredited calibration certificate
- 2 pcs Green terminal plug (For switch test)
- 10 pcs O-ring 3.5
- Additel ADT151 Pressure module for ADT7XX controller
Pressure module for ADT7XX controller
Order #: ADT151
Mfg #: ADT151
Availability: Call for Availability$3,493.00
Specification | ADT793 |
---|---|
Pressure Range | 0 to 15,000 psi (1,000 Bar) |
Precision[1] | 0.008%RD or 0.007%FS or 0.01%FS |
Accuracy[2] | 0.01%RD or 0.01%FS or 0.02%FS |
Control Stability[3] | < 0.005%FS, typically 0.003%FS |
Control Response Time[4] | < 30 Seconds |
Pressure Type[5] | Gauge, absolute |
Internal Pressure Control Modules | 1 |
External Pressure Control Modules | 1 |
Max Pressure Range of Internal Module | 0~15,000 psi (0~1,000 Bar) |
Min Pressure Range of Internal Module | -15~3,000 psi (-1~200 Bar) |
Min Range of External Module | -15~1,000 psi (-1~70 Bar) |
Range Switching Mode[6] | Fixed, auto |
Pump Source Type[7] | Built-in hydraulic pump, no external pressure source required. |
Control Mode | Fast, standard, custom |
Maximum Overshoot | < 1%FS |
Maximum Load Volume | < 10,000 psi: Max: 80 ml, 50 mL recommend 10,000~15,000 psi: Max: 60 ml, 35 mL recommend |
Contamination Prevention System (CPS) | Optional |
Pressure Port | 1/4 BSP M |
Built-in Reservoir Volume | 800 mL |
External Reservoir Extension | Optional |
Port Filter[8] | Support |
Media | Diethylhexyl Sebacate or deionized water |