AEMC Instruments MR526 AC/DC Current Probe
AC/DC Current Probe 100AAC 150ADC,10mV/A & 1000AAC,1400ADC
Order #: 1200.83
Mfg #: 1200.83
Special Price
$570.57
Regular Price
$600.60
AEMC Instruments MR526 AC/DC Current Probe
AC/DC Current Probe 100AAC 150ADC,10mV/A & 1000AAC,1400ADC
Order #: 1200.83
Mfg #: 1200.83
Special Price
$570.57
Regular Price
$600.60
Availability: Ships in 1-2 weeks
Product Highlights
The MR526 is a hall effect current probe with 2 ranges designed for DMM and Data loggers. The probe has a 5 ft (1.5m) lead terminated by 4mm safety banana plug and can measure up to 1000AAC and 1400ADC. The probe is powered by a 9V battery or by a standard 5V external power supply via a micro-USB connector (not supplied).
100AAC, 150ADC, 10mV/A & 1000AAC, 1400ADC, 1mV/A, 4mm Banana Plug Lead
- The jaw shape enables users to clamp on to cables or small busbars
- Powered by battery or standard external 5V power supply via micro-USB connector
- Measures up to 1000Aac and 1400Adc (model dependent)
- Automatic Zero DC reset function
- Auto Power Off enable/disable function
- Millivolt output compatible with most equipment and instruments
- Battery life up to 50 hours
- Safety rating 600V CATIII
Maximum Conductor Size | 1.53" (39mm) or 2 bus bar 1.96 x 0.19" (50 x 5mm) |
Current Range | 0.5 to 100AAC, 150ADC & 0.5 to 1000AAC, 1400ADC |
Output Signal | 10mV/A & 1mV/A via a 5ft (1.5m) lead with 4 mm safety banana plug |
Accuracy 50/60Hz | ≤1.5% @ 100A & ≤2.5% @ 800A |
Frequency Range | DC to 30kHz (-3dB) (depending of current value) |
Phase Shift At 50Hz | ≤2°@ 100A ≤1.5° @ 800A |
Load Impedance | >1MΩ and ≤100pF |
Overload | 3000ADC or 1000AAC continuous for < 1kHz |
Zero Adjust | Automatic on both ranges |
Power Supply | 9V alkaline Battery (NEDA 1060A, 6LR61) or 5V DC Micro-USB type B |
Battery Life | 50H typical |
Low Battery Indication | Green LED blinking |
Overload Indication | Red LED on when measurement > than selected range |
Dimensions (H x W x D) | 9.31 x 3.82 x 1.73" (236.5 x 97 x 44mm) |
Weight (with Battery) | 1.15lbs (520g) |
Operating Temperature | 14° to +131°F (-10 to +55°C) |
Storage Temperature | -40° to +176°F (-40 to +80°C) |
Relative Humidity | Up to 85% @ 35°C |
Ingress Protection (Ip) | IP 40 |
EMC | EN 61326-1 |
Safety Compliance | IEC 61010-1, EN 61010-2-32, Pollution Degree 2, 600V CAT III |